Optical three-dimensional metrology of wafer dicing grooves
Optical three-dimensional metrology of wafer dicing grooves
Author(s):
Affiliation(s):
Corresponding author and e-mail:
Abstract
Inspection of a diced wafer is difficult because a single groove contains optically dissimilar features: reflective lands, a recessed bottom, fractured walls and localized edge damage. Converting their signals into dimensions therefore depends on both visibility and calibration. This review considers coherence scanning interferometry, confocal and chromatic confocal microscopy, microscopic fringe projection and complementary optical inspection from that measurement perspective. For each representative study, we retain the measurement objective, acquisition and reconstruction route, reported quantitative result and stated operating constraint. Axial resolution, repeatability, dimensional uncertainty and defect detectability are treated separately and tied to the groove quantity they support, such as width, depth, wall angle or damage. The recurring failure at steep walls is traced to angular collection, opening-induced occlusion and responses that mix more than one height. These observations lead to a hybrid inspection concept in which rapid screening is followed by local dimensional measurement and, where needed, a change of viewing geometry. Three research directions are considered in detail: visibility-guided acquisition, joint calibration with propagated uncertainty, and tolerance-driven inspection on etched reference trenches and mechanically damaged grooves. They are assessed using dimensional error, valid surface coverage and complete inspection time. The aim is to connect optical-sensing capability with reproducible groove measurements and useful feedback to the dicing process.
Keywords: wafer dicing grooves; optical metrology; coherence scanning interferometry; chromatic confocal microscopy; structured light; hybrid inspection
Introduction
Dicing fixes the boundary of each die, but it can also damage the adjacent material and reduce the usable die area. Hamdy et al. [1] review chipping produced by diamond-blade cutting of Si and SiC and relate the observed damage to material and process conditions. A different part of the problem is street consumption: Feng et al. [2] developed a rolling-slitting process for manufacturing 0.048 mm diamond blades and evaluated spindle current, chipping and kerf width in SiC cutting experiments. Cut position is addressed separately by Chen et al. [3] through servo-control design. Taken together, these studies imply that an inspection system must distinguish at least three outputs: where the street lies, how much material has been removed, and how far the surrounding damage extends.
Laser-based separation changes the quantities that have to be inspected. Zhang et al. [4] study asynchronous dual-beam dicing of 4H-SiC, whereas Wang et al. [5] examine structural evolution and defect generation during ultrafast stealth dicing. Using a layered power strategy, Yang et al. [6] separate 508 µm SiC wafers and report approximately 1 µm cross-sectional roughness together with a crystallographic dependence of fracture quality. Open-groove depth, internal modification depth and the texture of a wall after separation therefore refer to different manufacturing stages. Illumination, reference-plane selection and the dimensional definition should be chosen only after that stage has been identified.
De Groot [7] describes interference microscopy for surface topography, including coherence localization, phase recovery and surface-dependent error. Li et al. [8] review chromatic confocal systems through the source, dispersive optics, spatial filtering and spectral detection, while Hu et al. [9] classify microscopic fringe-projection architectures and Feng et al. [10] compare calibration models experimentally. These studies establish how height information is formed and reconstructed. A diced groove adds a separate geometric problem: the land, edge, bottom and wall are not visible under the same conditions, and coordinate errors at those locations propagate differently into width, depth and wall angle.
This review is restricted to exposed dicing grooves. Methods aimed at buried interfaces or whole-wafer geometry are considered only when they add information that an exposed-groove measurement cannot provide. The discussion starts from measurand definitions and signal formation, then moves to representative instruments, sidewall-access failures and hybrid acquisition. Quantitative results are reported together with the specimen, metric and acquisition conditions used in the cited work. When signal is lost, the emphasis is on which requested dimension is affected and what change in acquisition or calibration could recover it. The experiments proposed in Section 8 follow the same measurement logic.
Measurement targets and challenges
Fig. 1 Groove measurands and damage features. (a) Cross-section defining top opening width, depth and wall angle relative to the wafer plane. (b) Top view showing edge chipping and a surface crack.
In Figure 1, nominal groove geometry is treated separately from damage around the cut. Blade dicing may leave saw marks, fractured facets and chipped edges. Top-opening width is therefore obtained from intact edge segments; excursions outside the fitted boundary are classified as chipping rather than folded into the kerf estimate. Including damaged edge fragments in the fit would mix two process responses and increase the apparent width. Depth uses a different construction: it is the distance, along the normal to the land datum, from the fitted land to a bottom that is actually resolved. A section crossing debris or a chipped corner is not equivalent to an intact section and should not be interpreted as one.
Consistent labels are also needed when measurements are compared across machines. Chang et al. [11] study the loss of chipping-prediction performance after transfer between dicing machines. Their ensemble meta-learning method combines hidden Markov, variational-autoencoder and bidirectional recurrent models and achieves 93.21% prediction accuracy in the reported cross-machine study. Because the method adapts to changing signal distributions, the result also exposes a metrology requirement: the definition of chipping and the dimensional labels must remain fixed, otherwise a change in the inspection pipeline can appear to be a change in the process.
The measurements of primary interest are kerf width, groove depth, sidewall angle and the extent of damage. Each quantity needs its own operational definition, reference coordinate system and rule for rejecting invalid observations. Table 1 lists those definitions together with suitable observation channels. Reports should also state whether width means the top opening, a section at a specified depth or a reconstructed cross-section, and whether depth is referenced to a visible bottom or to an internal interface.
Table 1. Key measurement indicators for wafer dicing grooves
| Quantity | Operational definition | Unit | Observation requirement |
| Top opening width | Distance between fitted undamaged opening edges in a stated section | µm | Calibrated lateral imaging or valid 3D data; exclude chipping consistently |
| Open-groove depth | Separation along the datum normal between fitted land and valid bottom | µm | CSI or confocal height; bottom must be observable |
| Wall angle | Inclination of a fitted wall relative to the wafer datum plane | ° | Visible wall span and fit interval must be specified |
| Bottom form | Departure of the valid bottom from a defined fitted surface | µm | State form-removal and exclusion rules |
| Surface texture | Specified profile or areal texture parameter after defined filtering | nm or µm | Report parameter, bandwidth, sampling and valid area |
| Chipping | Lateral extent, area and, if resolved, depth of edge damage | µm or µm² | Intensity imaging screens; depth requires calibrated 3D data |
| Surface or internal crack | Observed crack indication and extent for the chosen channel | µm | Channel-specific crack contrast; depth from resolved interfaces or sectional reference |
| Residual material | Thickness or extent of a specified unremoved layer | µm | Requires identified interfaces or independent sectional verification |
| Centerline deviation | Offset from the designed street in a common coordinate frame | µm | Registration and edge definition contribute uncertainty |
| Across-groove variation | Distribution of consistently defined measurements | Varies | State sampling locations, invalid regions and repeat count |
The definitions in Table 1 are meaningful only if the associated observation rules are followed. Chipping changes the edge available for fitting, debris can alter which point is taken as the bottom, and a short visible section of wall weakens an angle estimate. Friedrich et al. [12] augment visual-inspection data with simulated dicing faults when real defect examples are scarce. Their approach can improve defect localization, but a dimensional scale still requires geometric calibration. Using common cross-section coordinates and an unchanged damage definition then allows local observations to be combined at wafer scale. Table 2 summarizes the optical channels used for these different roles.
Table 2. Optical channels and their roles in groove inspection
| Channel | Measured observable | Inspection role | Controlling condition |
| CSI / WLI | Coherence envelope and fringe phase | Local depth, steps and accessible texture | Resolvable surface response; calibrated axial scan |
| Confocal / CCM | Axial intensity or spectral peak | Height and line profiles | Identifiable peak; field and displacement calibration |
| Microscopic FPP | Fringe phase and calibrated rays | Area geometry and street screening | Shared illumination and observation; valid phase order |
| Bright / dark field | Image contrast and scattering | Edge and defect localization | Calibrated lateral scale; representative defect labels |
| Spectral / backside | Wavelength-dependent optical path | Selected buried interfaces and residual layers | Transmission and identified layer model |
| Hybrid acquisition | Registered complementary observations | Feature-specific dimensions and damage assessment | Common datum, covariance and acquisition budget |
Optical return varies sharply within a single groove. Smooth lands tend to produce a strong specular signal, whereas fractured facets scatter light over a broader angular range; both can appear in the same field together with a weak wall return. A deep opening further limits illumination and collection. These mechanisms call for different remedies. Exposure control addresses dynamic range, signal estimation addresses noise in data that have already been acquired, and a new optical path is needed when the problem is occlusion. Diagnosing the mechanism first makes the choice of a complementary channel more specific.
The channels do not measure the same observable. Interferometric and confocal systems assign a local surface to an axial coordinate; fringe projection combines measured phase with calibrated viewing geometry; bright-field or dark-field images mainly show edges and scattering features. The outputs become complementary only after each is tied to a defined measurand. Image contrast, for example, can mark the boundary of a chipped region, but its depth still requires calibrated height data. The following sections compare the channels and trace the measurement chain that turns each observable into a groove quantity, including the conditions required for that chain to remain valid.
Coherence scanning interferometry
Principle and height estimation
Coherence scanning interferometry (CSI) records the interference response while the optical path difference is scanned. ISO 25178-604 [13] specifies the relevant instrument characteristics. Axial position is first localized by the coherence envelope, and phase can then refine the position within that response. The term white-light interferometry (WLI) is retained where it is used by the cited study. Montgomery et al. [14] provide an example in which CSI is combined with environmental chambers for functional-material characterization; in that arrangement, the specimen environment and the strength of the surface return both form part of the measurement conditions.
A broadband source illuminates the reference and sample paths. Their returning fields interfere at a detector, and scanning records an intensity trace at each lateral position. Figure 2 illustrates the relation between the optical paths and the localized interference envelope. A usable trace requires sufficient returned light, suitable path matching and a response that can be assigned to the surface of interest.
Fig. 2 Low-coherence interferometry. (a) Sample and reference paths. (b) Calculated single-surface response illustrating a localized coherence envelope and its carrier fringes; intensity is in arbitrary units.
Height recovery involves scan calibration, envelope detection and phase estimation. Ma et al. [15] use carrier-phase distributions to correct scan-step noise, targeting deviations of the axial sampling coordinate. Wu et al. [16] compare centroid, Fourier-transform and Hilbert-transform envelope detectors through simulation and step-height experiments. Centroiding is fast at low noise; Fourier and Hilbert processing improve noise resistance, with the Hilbert implementation faster than the Fourier implementation in their comparison. These results support choosing an estimator from the measured noise regime and the available processing time.
Equation (1) uses a Gaussian coherence envelope with a carrier fringe to represent a single reflector. The width parameter is expressed in scan-coordinate units, and the round-trip reflection path gives the carrier the factor of two. This approximation is suitable for a single surface viewed at approximately normal incidence in air. At abrupt edges or multilayer interfaces, diffraction and overlapping returns can alter the response enough that the reduced model no longer describes it adequately.
Here, B is the background intensity and A is the modulation amplitude. The parameter z₀ locates the envelope center, s sets its width, λ₀ is the central wavelength and φ₀ is the initial phase. The model assumes one reflecting surface, propagation in air and incidence that is approximately normal.
Noise can obscure the maximum of the coherence envelope. Li et al. [17] address this by applying complete ensemble empirical mode decomposition with adaptive noise before envelope localization, separating useful oscillatory content from the recorded signal; their simulations and experiments show improved peak recovery. The processing acts on an interferogram that has already been acquired, so it cannot by itself determine which of several possible heights produced an overlapping response. For groove data, peak width, asymmetry and secondary peaks should therefore be inspected together with the localized height. After background removal, Equation (2) expresses the centroid for non-negative envelope weights.
Here wᵢ is the extracted envelope weight at scan position zᵢ within the chosen response window. The centroid is meaningful only when that window is physically appropriate and the denominator is nonzero.
Li and Zhang [18] reduce the number of phase observations by using two-step orthogonalization. A differential signal is formed from two phase-shifted interferograms; filtering and normalization are followed by Lissajous ellipse fitting to recover phase for an arbitrary phase step. The reported simulations and experiments support this phase-extraction procedure. Fewer phase observations can lessen sensitivity to changes between frames, but a groove topography still depends on choosing the correct surface response and on calibrated spatial sampling.
Larkin [19] treats coherence-envelope localization and the subsequent phase refinement as separate steps in white-light interferometry. The distinction is useful at a groove transition because the two failures have different signatures: choosing the wrong envelope order gives a relatively large height jump, while phase noise perturbs a surface whose envelope has already been selected correctly. At land-to-wall and wall-to-bottom boundaries, examining these error modes separately helps determine whether an apparent improvement comes from better edge reconstruction or merely from smoothing the profile.
Acquisition architecture and access to difficult surfaces
The acquisition strategies in Fig. 3 address four different bottlenecks: optical-delay scanning, temporal phase sampling, lateral field coverage and access to recessed structures. Their engineering value is determined by the bottleneck that dominates the required groove measurement.
Kim et al. [20] target rapid profiling of surfaces with strongly varying reflectance. A fiber stretcher modulates optical delay, asymmetric line illumination extends the usable intensity range, and dispersion and polarization compensation stabilize the interferometric response. The reported axial resolution is 14.2 µm. For groove inspection, this architecture is relevant to rapid profile capture across mixed reflectance; the resolvable separation of axial responses and the uncertainty of an individual fitted height are separate quantities to evaluate.
Kahatapitiya et al. [21] review optical coherence tomography (OCT) for industrial inspection. OCT organizes low-coherence returns by depth and is useful for interfaces reached through a transmitting material. In dicing, that role concerns internal modification or remaining layers. Transmission, scattering and separation of neighboring interface responses determine the usable depth information. Surface CSI and depth-resolved OCT thus occupy different positions in the inspection sequence.
Bouchal et al. [22] use coherence-gated Mirau imaging for single-shot quantitative phase. The method reduces the need for sequential phase-stepped frames while retaining interferometric contrast and is suited to situations in which temporal disturbance limits acquisition. For an abrupt groove, the surface response and phase interpretation at edges remain the controlling spatial constraints. Figure 3 compares these acquisition functions.
Fig. 3 Interferometric acquisition strategies. Functional schematics show optical-delay modulation, coherence gating, lateral scanning and transmissive angular synthesis.
Zhao et al. [23] use epi-illumination spatio-temporal scanning interference microscopy (epi-STSIM) to enlarge the lateral field without sacrificing the local spatial resolution. During acquisition, the specimen is translated laterally and neighboring interferograms overlap. Their non-iterative generalized phase-shifting procedure combines Lissajous ellipse fitting with least-squares sine fitting so that vibration and stage-straightness errors in the phase step can be accommodated. For an etched circular feature, the reported mean depth is 202.67 nm with a standard deviation of 0.399 nm; WLI gives 203.93 nm and 0.764 nm. The two instruments therefore differ by approximately 1.3 nm for that feature.
Extending the field in the epi-STSIM experiment has a measurable time cost. The reported 4.274 × 0.204 mm area is built from 2000 frames acquired at 44 frames/s, corresponding to 45.45 s of capture, and a typical 2000-frame set takes approximately 2.9 min to process on the reported desktop computer. Reliable phase-step estimation depends on stable overlap and comparable reflectance statistics; complex material contrast, high roughness and densely inclined surfaces make that estimation harder. Epi-STSIM is therefore well suited to long streets with repeated accessible features, but fractured walls still need a separate check of phase-step stability across changes in slope and material.
Li et al. [24] use Fourier ptychographic CSI (FP-CSI) for a different access problem: a high-aspect-ratio trench observed through transmitting silicon. Near-infrared interferometry provides complex-field information, and angular-spectrum synthesis combines the angular views without iterative phase retrieval. The reported trench is 300 µm deep and 10 µm wide, an aspect ratio of 30:1. In ten trials, mean deviations from SEM are 1.00% for width and 0.24% for depth, with repeatability of 0.95% and 0.59%, respectively. These results show that the trench dimensions can be recovered in a configuration where direct reflective access to the bottom is severely limited.
The FP-CSI experiment uses 200 angular views. Capture takes approximately 9 min and reconstruction approximately 8 min. The specimens are transmitting silicon, particularly before metallization, and include deep-reactive-ion-etched trenches and multilayer MEMS structures. The clearest result is therefore access to top and bottom morphology in that transmissive configuration. A diced groove differs: its walls can be rough and strongly scattering, and metal layers can remove the transmission condition. Transfer to dicing should consequently be tested on matched etched and diced structures using bottom-depth error, wall coverage and acquisition time as separate outcomes.
The interferometric architectures above spend time in different places. Line-field and single-shot designs eliminate specific acquisition steps; epi-STSIM extends the lateral field while retaining local resolution; FP-CSI instead changes the optical path to reach recessed geometry. Selection should begin with the required measurand and then with the component of the inspection cycle that is limiting throughput, such as capture, computation, motion or a targeted repeat measurement.
Applicability and principal failure modes
CSI can provide local step height and texture only where the surface produces a usable coherent return. Petzing et al. [25] discuss rough-surface topography measurement in terms of both specimen response and processing choices. Chu et al. [26] apply white-light scanning interferometry to hybrid-bonding surfaces, a semiconductor problem in which the measured features are comparatively accessible. A diced groove is more restrictive because of its steep facets and narrow opening. Bottom visibility therefore limits depth measurement, while texture evaluation also depends on lateral transfer and the spatial bandwidth used for the calculation.
When both the land and the groove bottom are visible, depth can be measured from a locally fitted land plane and the bottom profile can be analyzed. Texture requires additional reporting choices. Form removal, spatial filtering, sampling interval and evaluated area all affect the result, as do debris, missing samples and interpolation. A reported sidewall roughness value should therefore be tied to the portion of wall that was actually resolved instead of being interpreted as a property of an unobserved complete wall.
Gao et al. [27] investigate slope-dependent CSI error using a cylindrical single-mode-fiber reference with low roughness and high roundness. A feed-forward neural network models the dependence on local tilt and is used to compensate the measured radius. This links a systematic error to a controlled reference geometry, but the resulting correction is still specific to the optical configuration and to the response of that reference surface. Rough, fractured groove walls would require calibration states that reproduce those conditions. Angular rejection and occlusion are considered separately in Section 7 because they determine which surface points exist in the data before any compensation can be applied.
Confocal and chromatic confocal metrology
Axial response and spectral encoding
Confocal sensing rejects much of the out-of-focus light by spatial filtering, and reflective height is obtained by locating a calibrated axial response. In a groove, however, the land, wall and bottom return light at different angles and intensities. The measurable region is therefore set by the combination of optical sectioning and geometric access, not by axial response alone.
A scanning confocal sensor localizes the surface using its focus-dependent response. Chromatic confocal microscopy (CCM) instead introduces longitudinal chromatic dispersion so that wavelength encodes focal position [8]. The measured spectral response is then mapped to displacement. This avoids a mechanical axial scan for each point within the calibrated range, although lateral scanning or multiplexing is still required for a surface map.
Figure 4 illustrates wavelength-to-height encoding. Source intensity, dispersive focusing, specimen reflectance and detector efficiency jointly shape the measured spectrum. The selected wavelength represents surface position after calibration of this complete response. At a layered edge, spectral weighting and contributions from neighboring heights alter the peak and become specimen-dependent error sources.
Fig. 4 Chromatic confocal depth encoding. Wavelength-dependent focal positions share an optical axis; the calibration curve illustrates the conversion from peak wavelength to displacement.
Liu et al. [28] use the strong dispersion of a phase Fresnel zone plate to extend the chromatic-confocal displacement range. Their fabricated 11.27 mm diameter element has a primary focal length of 50 mm. The calibrated axial range exceeds 16 mm, with a reported axial resolution of 0.8 µm and displacement measurement accuracy better than 0.4%. Step-height, glass-thickness and surface-profile measurements show the sensor functions. The design is range-oriented: the displacement span is obtained together with finite spectral discrimination and a specific calibrated optical response.
Liao et al. [29] increase both lateral efficiency and displacement range with an off-axis co-optical-path line sensor. The reported line length is 12.5 mm, and the calibrated axial range exceeds 20 mm over 450–750 nm. For specular samples, the allowable tilt is ±7°. That angular limit is important for groove work: extending the axial range does not remove the orientation restriction on a specular wall. A line crossing a groove edge must therefore be evaluated for field calibration and optical access as well as nominal range.
Two related developments act on different parts of the chromatic-confocal chain. Wang et al. [30] study a dual-optical-path, dual-grating line-confocal arrangement to manage the range–resolution trade-off, using both simulation and experiment. Chen et al. [31] use multispectral sensors to increase the rate of microscopic surface profiling. In the first case, the optical dispersion controls how displacement is encoded; in the second, the spectral detector controls how quickly that encoding is sampled. For line inspection, the useful rate is therefore the number of independently resolved profiles obtained across the calibrated field in a complete measurement cycle.
Focus measures and confidence in height recovery
Shape from focus (SFF) is another way to infer geometry from an axial response, although the observable is different from that of a confocal sensor. Conventional SFF evaluates image sharpness through an axial stack, whereas confocal measurement localizes an optically sectioned intensity or spectral peak. Both require a peak position to be selected. On a reflective land, however, image brightness varies with illumination and reflectance as well as focus, so axial position should be derived from a structural focus measure rather than brightness alone.
Eq. (3) places the SFF surface at the stack position where the chosen sharpness measure reaches its maximum. Gradient-, Laplacian- and frequency-based measures all evaluate a spatial neighborhood, so texture and noise influence the location and width of that maximum. On a weakly textured land or at a sharp groove boundary, the peak may shift or broaden; the resulting height can then be biased or poorly constrained even when one image in the stack appears to be the sharpest.
F denotes the local image-sharpness measure evaluated through the axial stack. The reliability of the recovered focus position depends on the spatial neighborhood used by F, its response to noise and the amount and type of image texture.
Gladines et al. [32] acquire SFF data during continuous motion in additive manufacturing rather than stopping at each plane. Camera triggering is synchronized with a tunable lens, and the moving image stack is registered before depth reconstruction. The cost of that motion compensation is evaluated against a stationary acquisition and a known reference. A similar strategy could reduce stop-and-settle time along wafer streets, provided that the registration preserves the groove-edge position despite local texture changes.
Autofocus can also fail because the focus metric itself is ambiguous. Weng [33] combines a Fourier structural metric with a convolutional-neural-network decision and evaluates local versus global peaks on Cu–Ni–Al alloy and plant-section images. Ashfaq and Mahmood [34] instead reweight focus volumes with a Difference-of-Gaussians-based perceptual factor, tested on fourteen synthetic and six real datasets. In both cases the objective is to select a peak supported by local image structure. For groove profiling, that selection still has to be converted to dimension by a physical displacement calibration.
A groove reconstruction should retain more than the selected height. Response amplitude, saturation, peak width, secondary peaks and fit residuals provide evidence about whether the chosen point is trustworthy. This is especially important at a land–wall boundary, where a sharp peak does not by itself identify the surface that produced it. Reference sections or a changed view can resolve that ambiguity. Keeping the raw response together with the selected peak also allows failures caused by low signal, mixed heights and displacement-calibration error to be distinguished.
Published architectures and their inspection roles
Yang et al. [35] replace spectrometer-rate acquisition in chromatic confocal microscopy with time-stretch detection. A stretched supercontinuum pulse is focused by a chromatic lens, and a high-speed photodiode with a 3 GHz digitizer converts the depth-encoded spectrum into a temporal signal. The reported system reaches up to 1 MHz A-scan acquisition and 5 Hz volumetric imaging in nude-mouse-skin experiments. Those rates describe the depth-signal readout itself. For reflective semiconductor grooves, valid-profile rate will also depend on returned intensity, spot size and the fraction of the surface that is optically accessible.
For free-form surfaces, Zhang et al. [36] pair a differential confocal response with precise positioning and report a composite uncertainty of 12.4 nm under the stated conditions. The differential curve gives a sensitive indicator near its calibrated position. At a groove edge, however, partial illumination, angular scattering and simultaneous contributions from different heights can alter that curve. A groove-specific reference spanning those conditions is therefore needed to establish the usable positioning interval and its contribution to depth uncertainty.
Shen et al. [37] improve weak-return reflectance confocal imaging by integrating a self-powered perovskite photodetector. The detector is reported to have 86.3% external quantum efficiency at 520 nm and 132 dB dynamic range; the imaging system reports 1.10 µm resolution at 4 frames/s. Wafer topography and photovoltaic grids were used in the reported demonstrations. This configuration acts on sensitivity and detector noise after light has reached the receiver. It can assist a weak but visible groove return, whereas a facet that is geometrically blocked requires a change in the optical path rather than a more sensitive detector.
Confocal systems improve performance through several different design choices. Line acquisition increases lateral coverage; spectral or temporal encoding shortens axial readout; differential or high-sensitivity detection improves localization of a returned signal. A groove instrument can combine these functions, but the calibration then has to cover field position, displacement, reflectance and slope together. The useful performance measure is not a nominal sensor specification alone, but the uncertainty in groove width, depth and wall angle over the part of the surface that is actually measurable.
Calibration and limits of transfer to grooves
Li et al. [8] show that chromatic-confocal performance is set jointly by source bandwidth, dispersive optics, conjugate apertures and spectral processing. Increasing displacement per unit wavelength extends the range but magnifies an error in wavelength localization. Reducing the aperture rejects more defocused light while also lowering the collected flux, and nonuniform source output or specimen reflectance can shift the effective focus wavelength. For a groove, axial range, point rate and height uncertainty must therefore be considered together for the reflectance and slope conditions expected in the measurement.
At a groove edge, the finite optical spot can contain both land and wall. Fitting that mixed response with a single peak assigns one height to two surfaces. A rough wall broadens the angular return, while a narrow opening reduces the light reaching the bottom. Quality data should record the failed response together with the reported dimension that it affects. Depth needs a resolvable bottom peak; wall angle needs a sufficiently long, valid wall segment for a stable fit.
A second channel is worthwhile only if it supplies information that the first channel lacks. More exposure or a more sensitive detector can recover a photon-limited signal or avoid saturation, but neither changes a blocked line of sight. Changing the illumination or observation direction changes angular access, and a transmitting wavelength can reach some buried interfaces. Selecting the corrective action from the observed failure mode avoids repeating a measurement under conditions that make the same surface inaccessible.
Axial steps alone are not enough to calibrate a confocal system for groove measurements. The calibration set should also contain lateral edges, tilted facets and representative reflectance contrasts; for a line sensor, checks are needed near both the center and the ends of the field. These references establish the coordinate mapping and the range of responses that can be accepted across the field. Remeasuring selected references after stage travel or refocusing then quantifies drift and any change introduced when the inspection channel is switched.
A practical sequence begins with a reference reflector to characterize wavelength-dependent throughput, followed by known displacements for the wavelength-to-height fit. Lands, rough facets and layered edges can then challenge the normalization and peak-selection rules under more representative conditions. Residuals should be inspected as functions of both field position and displacement, while independent check points are kept out of the fit. This separates spectral weighting, interpolation error and specimen-dependent response in the uncertainty budget instead of absorbing them into a single residual.
Structured light and microscopic fringe projection
Phase recovery and geometric calibration
Fringe projection profilometry (FPP) derives three-dimensional shape from the deformation of a projected pattern viewed by a calibrated camera-projector pair [10]. Reconstruction involves estimating phase, resolving phase or fringe order and converting the result with the calibrated geometry. FPP observes an area rather than a single point, which makes it attractive for rapid screening. The practical inspection rate, however, is set by the number of projected patterns, their exposure time and the proportion of pixels for which the phase can be reconstructed validly.
Zhang et al. [38] shorten the projected sequence with two-plus-two phase-shifted coding, so phase and coding information are retained with fewer frames. This is useful when specimen motion constrains the available acquisition time. It does not eliminate the local failure at a groove boundary: weak modulation, a geometric discontinuity and mixed pixels can still cause the wrong fringe order to be selected.
Measured fringe phase becomes geometry only after camera and projector coordinates have been calibrated. Figure 5 shows two common representations of that conversion: a phase-to-height mapping and a model based on calibrated rays. Which model is appropriate depends on the optical architecture, while the overlap of illuminated and observed regions determines where a valid phase measurement exists in the first place.
Fig. 5 Microscopic fringe projection. (a) Joint illumination and observation of a surface point. (b) Phase-to-height and calibrated-ray routes from measured fringe phase to three-dimensional coordinates.
Hu et al. [9] separate microscopic FPP systems based on stereomicroscopes from those using individual long-working-distance lenses. The stereomicroscope packages two optical paths in one compact assembly; separate lenses allow more freedom in working distance and the projection-observation angle. Telecentric imaging controls magnification variation but changes the calibration to an affine or telecentric form. Optical architecture and calibration are therefore linked decisions. Wang et al. [39] consider the same issue for small components with complex geometry. In a recessed groove, a larger triangulation angle increases geometric sensitivity but also reduces the surface region seen by both paths.
Short acquisitions can be achieved either by putting more information into a few projected patterns or by reconstructing from a single observation. Zeng et al. [40] use composite fringes with a multiperiod spatially varying phase shift and recover absolute phase from four patterns while limiting noise amplification. Wang et al. [41] use a super-resolved learned reconstruction and report three-dimensional imaging at 100,000 frames/s. The two results solve different information problems, so frame count alone is not a meaningful comparison. For dicing grooves, fringe sampling, boundary bias and processing latency still need to be measured on the material and geometry of interest.
Calibration volume and uncertainty in microscopic reconstruction
Feng et al. [10] experimentally compare linear, inverse-linear, polynomial and governing-equation phase-to-height mappings with triangular stereo calibration. Known-height planes are used for calibration, and a complex plaster object tests reconstructed shape. With dense sampling of the calibration volume, the inverse-linear and polynomial mappings perform well in their experiments. When points lie outside, or the volume is incompletely sampled, triangular stereo gives the best result of the tested models. The number of phase shifts is also varied. The study therefore ties model choice to calibration coverage: local flexible mappings are effective where data support them, whereas the geometry-constrained ray model is less dependent on local interpolation when extrapolation is required in the reported tests.
For a groove measurement, the calibration planes should cover the land, the deepest bottom and lateral positions near the edge of the field. A separate reference containing a known discontinuity can then test edge location and fringe-order assignment. The planar target tests the coordinate transformation; the discontinuous target tests the combined effects of sampling, visibility and phase recovery. Treating these tests separately makes it easier to distinguish a systematic calibration error from a boundary-specific reconstruction failure when width and depth uncertainty are evaluated.
Visibility and inspection suitability
Lv and Qian [42] formulate a precision model that links camera electronics and intensity statistics to fringe intensity, phase and reconstructed geometry. It can be used during design to select exposure and geometry, or on an existing instrument to estimate how precision varies across the field. For a groove, this provides a way to propagate local measurement noise. Bias in the calibration, an incorrect fringe order and occlusion still require separate checks because they break the assumptions of the locally valid phase model rather than simply increasing its noise level.
Meng et al. [43] extend the depth of field of microscopic FPP with a coaxial light-field arrangement. The coaxial layout changes the balance between projection and observation, while the light-field data support the depth reconstruction. In a narrow trench, the relevant geometric question is whether a surface point is both illuminated and observed at its depth. Computing that overlap identifies the wall and bottom regions for which the system actually provides a usable constraint.
Zhang et al. [44] reduce blind regions from shadowing, occlusion and specular reflection by combining a vertical telecentric projector with four oblique tilt-shift cameras. Corrected projection modeling and bundle adjustment register the views to one coordinate frame. Over a 70 × 40 × 20 mm measurement volume, the reported repeated-measurement error and standard deviation do not exceed 10 µm. The multiple views improve coverage of complex surfaces. A microscopic groove would require the visibility analysis to be repeated at its smaller scale and narrower opening, although the common-calibration strategy can still be used.
Wang et al. [45] analyze harmonic crosstalk in microscopic temporal frequency-division multiplexing. Their guard bands reserve spectral space for second harmonics so that those components do not overlap the fundamental channels before phase demodulation. Experiments on discontinuous planar and spherical surfaces support the reconstruction. The method addresses a coding-related phase error; saturation and shadowing arise from different mechanisms and are better handled by exposure control or view selection. A groove inspection record should therefore preserve the cause assigned to each failed region rather than treating all invalid phase as the same failure.
Microscopic FPP is most useful for area geometry where projected fringes retain sufficient modulation and both illumination and observation reach the surface. In a hybrid system, the FPP map can first screen a street and nominate locations for more precise measurement. CSI or confocal sensing can then be reserved for local depth and texture. The value of that division is not the screening rate alone: it should reduce the number of expensive local measurements without losing the dimensional excursions and defects that matter to process control.
Hybrid metrology and system integration
A hybrid system is useful only when its channels contribute genuinely different constraints to the same defined quantity. In wafer-groove inspection, this may mean combining wide-area localization with local height, axial positioning with defect contrast, or physical thickness with optical path length. Because each pairing has a different calibration structure, the channels need a common datum, their own quality indicators and an uncertainty model that carries shared as well as channel-specific terms.
Chen et al. [46] use OpticFusion to combine multi-view WLI geometry with optical-microscope appearance in a neural implicit reconstruction. Height-related observations from the interferometric views constrain shape, while microscope images provide texture. The combination is informative at the microstructure scale, but a dimensional application requires the contributions of measured height, image texture and regularization to be checked separately. Particular attention is needed near view boundaries and in regions constrained by only a small number of observations.
That separation suggests a conservative fusion strategy. Dimensions should be fitted mainly from calibrated geometric observations, while texture is used to locate features or to assign material and damage classes. The reconstruction should retain the origin of each geometric constraint. Independent edge and step references can then show whether regularization has displaced a boundary. Only after those shifts are quantified can a textured reconstruction be treated as a dimensional result rather than a visualization.
Ye et al. [47] measure wafer bow and warp with a full-field interferometric arrangement in which an optical flat forms an air gap above the wafer and a large-aperture collimated beam generates fringes over the field. Relative bow and warp deviations are reported below 3% when the result is compared with a commercial interferometer. Such a global surface map changes slowly across the wafer and can therefore serve as a reference for local focus planning and coordinate initialization.
The global wafer form can be used to reduce the search required by a local groove sensor. At each street, the measured surface predicts a narrower axial interval for focus or scanning; the remaining form and alignment error can then be removed by a local datum fit before depth is calculated. The benefit should be tested directly by comparing positioning time and local dimensional uncertainty with and without the global-form correction.
Hu et al. [48] separate focusing from defect contrast by combining laser-confocal positioning with dark-field imaging. Their reported result is detection of 55 nm patterned defects over a 4 mm line range, where the signal is the scattering produced by those defects. For a dicing groove, the same division of functions could be used to locate debris, chips or crack indications rapidly and then send selected sites to a calibrated geometric sensor. The decision threshold would still need reference defects of known morphology so that dark-field contrast can be connected to a detection criterion.
Backside access through silicon is treated by Choi et al. [49] with infrared chromatic-confocal sensing combined with spectral interferometry. For a nominal 534.2 µm silicon wafer, they report thickness repeatability of 72.3 nm and combined standard uncertainty of 560.7 nm (k = 1), together with approximately 14 µm spatial resolution. The study also includes backside depth measurement of Cu-filled through-silicon vias using a high-NA spectral-interferometric configuration. This is relevant to groove inspection when a transmitting path to a buried interface is required; an exposed opaque groove wall still requires direct top-side access.
The same wafer-thickness experiment reports 534.05 µm for a nominal 534.2 µm wafer and 72.3 nm repeatability over 20 measurements. The chromatic-confocal and interferometric contribution to the uncertainty is given as 556.7 nm, while inclusion of wafer flatness and thermal expansion gives a total standard uncertainty of 560.7 nm. Spatial resolution is approximately 14 µm. The contrast between the small repeated-measurement spread and the larger total uncertainty shows that optical calibration and specimen terms remain important even when the readout is stable. The Cu-filled TSV result is obtained with its own spectral-interferometric configuration.
The hybrid thickness measurement uses two independent observables: chromatic focal separation and interferometric optical thickness. Combining them permits physical thickness to be estimated without assuming that refractive index is known from a single signal. The reported per-point timing is approximately 0.5 s for spectrum acquisition and 0.3 s for processing, or about 75 points/min before stage motion. At that rate, the method is better matched to selected checks of residual layers or buried interfaces than to dense dicing-street mapping. The layer sequence, refractive index and spatial sampling therefore have to be specified in the model used at each selected location.
Zhang et al. [50] use BWD-DETR for bright-field defect detection on textured wafer backgrounds, where defect contrast is weak. Their wavelet backbone and feature-fusion modules give reported values of AP50 = 96.56% and AP50:95 = 54.94% on the study dataset. The lower value at the stricter overlap range indicates that localization quality becomes more demanding as the criterion tightens. In a hybrid metrology chain, such a detector is best treated as a source of candidate coordinates; a calibrated mapping transfers those coordinates to the dimensional sensor, and held-out process conditions are needed to quantify missed defects and false alarms.
Park et al. [51] combine microsphere-assisted imaging with hyperspectral information and report 66 nm optical resolution within a 5.6 × 5.6 µm field. The method couples fine spatial detail to wavelength-dependent response, but the field is small enough that it is better suited to local verification than to wafer-scale screening. Applying it to a dicing groove would require access to the selected region and, whenever spectral information is interpreted as thickness or composition, an explicit material model.
The sequence in Figure 6 begins with a wafer-coordinate screening map, which identifies street segments and possible defects for local measurement. The selected local sensor then acquires the dimensional signal and evaluates its quality. If the observation fails, the corrective action is chosen from the diagnosed cause: exposure is changed for an intensity problem, the view is changed for an access problem, and a transmitting wavelength is used when a buried interface is the target. Valid observations are registered before width, depth or wall angle is fitted. Dimensions, uncertainty and feature-specific coverage are carried together so that any additional acquisition is directed at the part of the requirement that remains unmet.
Fig. 6 Proposed hybrid groove inspection with feature-specific quality evaluation and targeted acquisition. The selected action follows the diagnosed source of missing or uncertain information.
Registration and measurement decisions
Gollee et al. [52] calibrate a multisensor scanstation containing blue laser line scanners, a three-axis motion unit and a turntable. Their common extrinsic calibration reduces the mean point-to-plane distance from 1.403 mm to 0.054 mm. Validation with an independent measuring sphere gives a probing size error of 0.042 mm and a probing dispersion of 0.248 mm. For a microscopic groove system, the relevant lesson is the separation of two questions: how well the calibration fits the data used to estimate it, and how well the calibrated system measures an independent geometry.
At microscopic scale, the shared fiducials must cover both the lateral field and the height range so that translation, rotation and scale can be estimated. A subset should be excluded from fitting and measured again after stage motion, refocusing or a channel change. Those repeated, withheld points measure drift at the inspection location rather than only the residual of the original fit. Parallel streets also need unambiguous identifiers to prevent registration to an adjacent street. Once the transformation is established, its coordinate covariance can be carried through the datum fit and the extracted cross-section.
Weights used in the fused fit should represent correlation as well as uncertainty. Errors from the stage, datum and common reference act on several channels and therefore remain present after fusion. They belong in the estimate of the common geometry, while residuals specific to each channel can be used to identify disagreement between sensors. Image contrast may determine where a candidate feature lies, but the dimensional fit should include only observations that satisfy the stated definition of the edge, bottom or wall.
Quality acceptance is feature-specific. Depth requires a valid datum and bottom; wall angle also requires a sufficient wall span and well-constrained lateral coordinates. A narrow missing wall patch has different consequences for those two outputs. Reporting coverage separately for the land, bottom and each wall links the quality decision to the dimension being delivered and identifies the next useful observation.
Steep sidewall signal loss and dimensional uncertainty
Steep-wall measurement failure combines angular rejection, geometric occlusion and surface-dependent signal distortion. Estimation and registration convert these effects into missing points or biased dimensions. Figure 7 separates the ideal specular collection constraint from the opening-limited angular window. These reduced geometries identify which part of the optical path needs to change before more precise signal processing becomes useful.
Fig. 7 Optical access to recessed features. (a) A normally incident collimated ray is deflected through 2θ by a facet tilt θ, relative to a collection cone of half-angle α. (b) The bottom-center opening has half-angle β = atan[w/(2d)]. (c) Calculated β against aspect ratio d/w, with 10:1 and 30:1 examples.
For an objective with numerical aperture NA in refractive index n, the collection half-angle is α = asin(NA/n). Consider first the restricted case of collimated illumination at normal incidence. A smooth facet tilted by θ from the wafer plane sends the specular return through an angle 2θ, and the collection condition is therefore 2|θ| ≤ α. De Groot [7] notes that pupil-filled illumination introduces oblique incident rays, shifting the ideal specular tilt limit toward α. In both cases, the usable signal falls as the overlap between illumination and collection pupils becomes smaller. Scattering can make steeper slopes visible, but it also changes the spatial response; recovery of form and transfer of fine texture should therefore be evaluated separately.
The groove opening imposes a second angular restriction. At the center of a straight trench bottom, the two opening edges define a clear half-window β = arctan[w/(2d)], where w is opening width and d is depth. Away from the center, or for tapered walls, the two limits are no longer equal and visibility has to be calculated locally. Once the specimen opening becomes the limiting aperture, increasing the nominal objective NA mainly adds rays that are blocked before reaching or leaving the surface. The available angular range is thus set by the objective and the local groove geometry together.
The scale of the restriction is illustrated in Fig. 7. For a trench 300 µm deep and 10 µm wide, β ≈ 0.95°; at the same depth, a 30 µm opening increases β to ≈ 2.86°. Even the wider opening leaves a narrow reflective window at these aspect ratios. FP-CSI avoids this specific top-side path by operating through transmitting silicon [24]. Numerical aperture by itself is therefore not enough to describe access; wavelength and viewing geometry can be just as important.
A larger pupil also changes the character of the recorded response. It admits more scattered directions and can mix more angular contributions into one spatial sample. In CSI, diffraction at edges and multiple scattering alter fringe visibility and the recovered height. In confocal measurements, contributions from neighboring heights can broaden or split the axial peak. In FPP, reduced modulation or a discontinuity can make fringe order ambiguous. Retaining these channel-specific quality measures allows the subsequent corrective action to be matched to the physical failure instead of to a generic low-quality flag.
Consequences for depth and wall fitting
Missing data are concentrated rather than random: steep or weakly returning parts of the wall disappear first. A wall fit based only on the surviving points can therefore overrepresent the optically accessible facets and shift the estimated angle. At the bottom, edge light can pull the reconstructed height toward the land, while ringing may introduce false extrema. Filling these gaps by interpolation creates a continuous-looking surface but does not restore the lost optical information. The uncertainty assigned to the requested dimension should consequently reflect the valid fitting span and the sensitivity to where its boundaries are placed.
Coverage can be reported explicitly rather than hidden inside a fitted surface. One useful quantity is the valid-area fraction in a predefined region of interest (ROI); for a uniformly sampled height map, Eq. (4) expresses the corresponding missing fraction. Both the ROI and the validity threshold must be fixed before instruments are compared. If the points are not uniformly distributed, an area-weighted calculation or a common evaluation grid is needed, because raw point count can favor dense sampling of the easy regions. Here the numerator is the number of valid pixels and the denominator is the total number of pixels in the ROI.
Depth is referenced to a fitted datum only after the relevant coordinate frames have been aligned. In the reduced case where top and bottom heights are estimated in the same frame, d = zₜ − zᵦ and Eq. (5) gives the corresponding uncertainty relation. A shared calibration or drift term introduces covariance between the two heights, so that term cannot be omitted. Errors from datum fitting, lateral position, registration and model mismatch must either be included in the height estimates or propagated through the complete measurement model.
Reporting a wall angle requires more than the fitted slope. The length of wall that was actually valid, the fit residuals and the lateral and axial coordinate uncertainties are also relevant. Repeated scans quantify short-term variability, but they can repeat the same geometric bias. An independent section or a second view can expose that bias. The difference matters most for a near-vertical wall, where a small lateral coordinate error produces a comparatively large change in the fitted inclination.
For a wall segment with vertical extent H and lateral run L, the angle to the wafer plane is θ = atan2(H, L). First-order propagation gives δθ ≈ (L δH − H δL)/(H² + L²) radians; close to a vertical wall this reduces to δθ ≈ −δL/H. For example, 0.5 µm of lateral error over a 100 µm valid span contributes approximately 0.29°, while the same error over 20 µm contributes approximately 1.43°. Increasing the valid wall span and improving lateral registration can therefore reduce angle uncertainty even when axial repeatability does not change.
Matching the remedy to the failure
The corrective measurement should be chosen from the observed failure. Saturation or a weak but visible return points to exposure or sensitivity; angular loss points to a different illumination or viewing direction. If the substrate transmits at an appropriate wavelength, a backside or alternative-wavelength measurement may reach an otherwise hidden interface. A model-based correction is appropriate only inside the response range for which it has been calibrated. After the change, the relevant feature-level quality metric should show whether depth, width or angle uncertainty has actually been reduced enough for the inspection decision.
Comparative assessment and research directions
Table 2 assigns inspection roles to the optical channels, and Table 3 compares representative numerical results. The studies optimize different parts of the measurement chain: axial response separation, repeated height localization, spatial resolution, defect detection or calibration. Their numerical values become useful selection criteria when matched to the required groove quantity, optical access and inspection time.
Table 3. Representative quantitative results with metric definitions
| Study and channel | Reported performance | Specimen or acquisition | Operating conditions |
| Kim et al. [20]Line-field WLI | Axial resolution: 14.2 µm | Fiber-stretcher delay modulation; asymmetric line illumination | Mixed-reflectance profiling; dispersion and polarization compensation |
| Zhao et al. [23]Epi-STSIM | Etched-circle depth: 202.67 nmStandard deviation: 0.399 nm | Wide-field demonstration: 4.274 × 0.204 mm2000 frames in 45.45 s | Depth and field results from separate demonstrations; ≈2.9 min processing; stable overlap and reflectance required |
| Li et al. [24]FP-CSI | Width / depth deviation from SEM: 1.00% / 0.24%Repeatability: 0.95% / 0.59% | 300 µm depth; 10 µm width; 10 trialsCapture ≈9 min; reconstruction ≈8 min | Etched transmitting silicon; 200 angular views; top and bottom morphology |
| Liu et al. [28]Fresnel-zone-plate CCM | Axial resolution: 0.8 µmDisplacement accuracy: better than 0.4% | Calibrated axial range >16 mm | Fabricated diffractive probe; step height, glass thickness and surface profiles |
| Liao et al. [29]Line CCM | Reported axial resolution: ±1 µmSpecular tilt range: ±7° | 12.5 mm line; axial range >20 mm450–750 nm band | Off-axis common optical path; line-wise spectral calibration |
| Yang et al. [35]Time-stretch CCM | A-scan rate: up to 1 MHzVolumetric imaging: 5 Hz | Multi-depth imaging of mouse skin | Time-encoded spectrum; reflective-groove response requires material-specific evaluation |
| Zhang et al. [36]Differential confocal | Composite uncertainty: 12.4 nm | Free-form-surface positioning | Calibrated differential response; edge mixing and slope govern groove transfer |
| Shen et al. [37]Sensitive confocal | System resolution: 1.10 µmImaging: 4 frames/s | Wafer, tissue and photovoltaic-grid demonstrations | Perovskite detector; 132 dB detector dynamic range |
| Wang et al. [41]Single-shot FPP | 3D imaging: 100,000 frames/s | Learned single-shot reconstruction | Dynamic imaging; fringe sampling and training-domain dependence |
| Zhang et al. [44]Multiview FPP | Repeated-measurement error and standard deviation: ≤10 µm | 70 × 40 × 20 mm volume | Telecentric projection; four cameras; joint calibration |
Table 3. Representative quantitative results with metric definitions (continued)
| Study and channel | Reported performance | Specimen or acquisition | Operating conditions |
| Ye et al. [47]Wafer interferometry | Relative bow / warp deviation: <3% | Global wafer form | Comparison with commercial interferometer; useful for local focus planning |
| Hu et al. [48]Confocal + dark field | Detected patterned-defect size: 55 nm | 4 mm line range | Scattering-based detectability with confocal focus positioning |
| Choi et al. [49]Backside hybrid | Repeatability: 72.3 nmTotal standard uncertainty: 560.7 nm | 534.2 µm nominal wafer; 20 repeats≈0.8 s/point before stage travel | Optical subtotal 556.7 nm; total includes wafer terms; ≈14 µm spatial resolution |
| Park et al. [51]Microsphere hyperspectral | Optical resolution: 66 nm | 5.6 × 5.6 µm field | Local semiconductor inspection with spatial and spectral information |
| Gollee et al. [52]Multisensor platform | Mean point-to-plane distance after calibration: 0.054 mmSphere diameter error: 0.042 mmProbing dispersion: 0.248 mm | 34 ± 0.002 mm measuring sphere; probing dispersion evaluated after excluding the 5% most distant points | Macroscopic platform; calibration strategy and independent validation inform system integration |
Figure 8 groups the reported dimensional-scale values by what they actually represent. Although the numerical values span several orders of magnitude, they refer to different metrics, including resolution, repeatability, uncertainty, detectability and calibration error, and cannot be interpreted as one common accuracy scale. The specimen and acquisition conditions must remain attached to each value. Percentage-, rate- and time-based results are left in Table 3 because they cannot be placed on the same length axis. The literature therefore shows a separation between strong performance on a local sensing task and the time needed to inspect a complete groove. Fast temporal readout shortens signal capture, angular synthesis can recover recessed structure at a substantial acquisition cost, and backside spectral sensing adds access to selected interfaces in transmitting silicon. A hybrid design should use those channels selectively and judge the result by required dimensions, valid coverage, uncertainty and total inspection time.
Fig. 8 Quantitative evidence map of dimensional-scale metrics reported by representative studies in Table 3. Values are shown on a logarithmic length axis (µm) and grouped by metric type: resolution, repeatability/standard deviation, uncertainty, detectability, and calibration/dimensional error. Specimen, optical configuration, and evaluation protocol differ among studies; the figure therefore maps reported evidence rather than ranking instrument accuracy. Percentage-, rate-, and time-based results remain in Table 3 because they are not commensurate with a length axis.
The proposed research program combines visibility prediction, uncertainty-aware registration and process-oriented evaluation. Each direction has a defined hypothesis, measurement sequence and comparison. Table 4 specifies the principal controls. Etched references establish geometric response under controlled conditions, and real diced specimens add the chipping, debris and fractured facets encountered in production.
Table 4. Experimental controls for the proposed hybrid inspection program
| Comparison | Controlled conditions | Primary evaluation |
| Fixed top view / preset views / adaptive views | Same specimens, validity thresholds and total time budget | Wall and bottom coverage; depth and angle deviations; added cycle time |
| Independent / joint calibration | Same fiducials; held-out checks; identical stage travel | Check-feature deviations; transformation covariance; drift |
| Independent weighting / covariance-aware fusion | Same valid observations and shared datum | Dimensional error; uncertainty-interval coverage; angle sensitivity |
| Fixed precision scan / selective acquisition | Same held-out wafers and defect-recall target | Missed defects; false alarms; complete inspection time |
| Etched references / diced grooves | Matched dimensional scale; documented reflectance and damage | Response failure versus slope, aspect ratio and morphology |
| Optical result / registered reference section | Known section location; reference and preparation uncertainty | Signed width, depth and angle deviations; feature-specific coverage |
Visibility-guided acquisition for steep walls
One research direction is to choose each additional view from a local visibility calculation rather than from a fixed angular sequence. A screening profile supplies the initial groove geometry, and the illumination and collection pupils are already calibrated. Ray intersections with the estimated opening can then identify views that reach a wall or bottom region missing from the first measurement. Fringe modulation, spectral peak quality or confocal amplitude provides a second test that separates geometric occlusion from a weak return. The experiment should determine whether this diagnosis-based choice recovers more valid wall span for each added measurement than uniformly spaced views.
The comparison can be made on trenches that vary in aspect ratio, wall taper and surface finish. For every specimen, use the same total time budget to acquire a fixed top view, a preset multiview sequence and the adaptive sequence. Apply one validity threshold to all three cases and calculate coverage separately for the left wall, right wall and bottom. Reference sections provide depth and angle deviations, while the log of stage motion and acquisition gives the extra time required by each strategy. The key outcome is whether the selected views add dimensional information in the difficult regions, rather than simply adding points to land areas that were already visible.
Joint calibration and covariance-aware fusion
A second direction is to calibrate sensor geometry, stage motion and the wafer datum in one model. The reference artifact should place lateral fiducials, known height steps and tilted facets across the working volume. Sensor transformations and stage terms are estimated together, while a separate set of features is withheld for validation. Repeating those validation measurements after stage travel and after changing channels provides an empirical measure of drift. The comparison of interest is then between this joint model, which retains shared errors, and channels that are calibrated independently and aligned only afterward with a rigid point-cloud transformation.
For a reported quantity q, the calibrated-state contribution can be propagated as u²(q) = JΣJᵀ, with J as the sensitivity row vector and Σ containing coordinate, calibration and shared-datum covariances. Noise unique to one sensor is kept as a channel-specific term, whereas stage and reference errors that affect several channels appear as correlated terms. The raw observations can then be fused once with this covariance model and once under an independence assumption. Signed dimensional deviations, uncertainty-interval coverage on held-out artifacts and the change in wall angle after an imposed lateral misregistration provide direct tests of which weighting scheme better matches the measured metrological behavior.
Tolerance-driven scheduling and process feedback
The third direction is to schedule measurements by their expected contribution to an inspection decision. For each region, define the required width, depth or angle tolerance and a minimum feature coverage. Candidate actions include a longer exposure, an additional view or a different sensor. Rank them by expected reduction in the relevant dimensional variance per added cycle time, using empirical response models from the benchmark. Impose a coverage requirement so that low predicted variance on a small easy region cannot terminate a wall measurement prematurely. Stop acquisition when the dimensional and coverage requirements are met, or route the region to reference inspection when the available actions are exhausted.
Evaluate scheduling on real diced streets from varied process conditions, with held-out wafers reserved for testing. Compare fixed full-coverage precision scanning with screening followed by selective measurement at the same defect-recall target. Record missed defects, false alarms, dimensional deviations and the complete time for motion, focus, acquisition, processing and reacquisition. Associate these outputs with blade, feed and cut-position records using a common street coordinate. This enables a process-feedback study in which measured changes in chipping, depth and centerline position guide adjustment of the corresponding process variables.
The benchmark should move from controlled references to the actual process surface in stages. First establish displacement and lateral scale on traceable references; next use etched trenches to isolate the effects of aspect ratio and slope. Diced grooves are measured only after those checks, with material condition, debris and edge damage recorded for each specimen. All non-contact measurements should precede destructive sectioning, and the section locations must be registered to the optical maps. The comparison should include reference uncertainty, preparation effects and between-specimen variation. Publishing the raw responses, acquisition settings, calibration data and feature-level validity masks alongside the reported dimensions would allow the benchmark to be reproduced.
Conclusions
Three factors repeatedly limit optical measurement of a wafer dicing groove: whether the requested surface can be seen, whether the detected signal can be assigned to the correct surface, and how that signal is tied to a calibrated coordinate system. CSI and confocal techniques provide local axial information when a separable surface response is available, while microscopic FPP supplies area geometry only where projection and observation overlap. Scattering-based and spectral measurements add different information, such as defect contrast or access to selected interfaces. The literature shows useful performance in each role, but under different optical constraints and acquisition costs.
The next step is therefore not to replace these methods with one universal sensor, but to decide which additional observation is needed for the requested dimension. A visibility model can nominate a new view when wall or bottom coverage is missing; a joint calibration can carry shared coordinate error into width, depth and angle; and a scheduling rule can stop or extend acquisition according to the remaining uncertainty and coverage requirement. Testing those ideas on common etched and diced specimens, together with independent dimensional references, would make instrument choices directly comparable and would connect the optical result to dicing-process feedback.
Acknowledgment
Funding Data
Conflict of Interest
Data Availability Statement
All quantitative material discussed in this review is drawn from the cited literature; the review does not introduce a new experimental dataset.
Nomenclature
| Abbrev. | Meaning | Abbrev. | Meaning |
| CCM | Chromatic confocal microscopy | OCT | Optical coherence tomography |
| CSI | Coherence scanning interferometry | RMSE | Root-mean-square error |
| Epi-STSIM | Epi-illumination spatio-temporal scanning interference microscopy | ROI | Region of interest |
| FP-CSI | Fourier ptychographic CSI | SEM | Scanning electron microscopy |
| FPP | Fringe projection profilometry | SFF | Shape from focus |
| MEMS | Microelectromechanical systems | TSV | Through-silicon via |
| NA | Numerical aperture | WLI | White-light interferometry |
References
[1] Hamdy, K., Okunkova, A. A., Volosova, M. A., Ali, S., Ibrahim, A. M. M., Lee, H. P., and Grigoriev, S. N., 2025, “Chipping size in Si and SiC wafers dicing with a diamond saw blade – A review,” Journal of Materials Research and Technology, 39, pp. 5788–5799. https://doi.org/10.1016/j.jmrt.2025.10.134.
[2] Feng, Y., Li, K., Dou, Z., Zhang, Z., and Guo, B., 2022, “High-Speed Dicing of SiC Wafers with 0.048 mm Diamond Blades via Rolling-Slitting,” Materials, 15(22), 8083. https://doi.org/10.3390/ma15228083.
[3] Chen, F., Huang, J., and Xu, J., 2024, “High-precision servo control design and optimization for dicing semiconductor wafer,” Journal of Manufacturing Processes, 117, pp. 346–354. https://doi.org/10.1016/j.jmapro.2024.03.033.
[4] Zhang, Z., Wen, Z., Shi, H., Song, Q., Xu, Z., Li, M., Hou, Y., and Zhang, Z., 2021, “Dual Laser Beam Asynchronous Dicing of 4H-SiC Wafer,” Micromachines, 12(11), 1331. https://doi.org/10.3390/mi12111331.
[5] Wang, L., Zhang, C., Liu, F., Zheng, H., and Cheng, G. J., 2022, “Ultrafast Pulsed Laser Stealth Dicing of 4H-SiC Wafer: Structure Evolution and Defect Generation,” Journal of Manufacturing Processes, 81, pp. 562–570. https://doi.org/10.1016/j.jmapro.2022.06.064.
[6] Yang, B., Wang, H., Peng, S., and Cao, Q., 2022, “Precision Layered Stealth Dicing of SiC Wafers by Ultrafast Lasers,” Micromachines, 13(7), 1011. https://doi.org/10.3390/mi13071011.
[7] de Groot, P., 2015, “Principles of Interference Microscopy for the Measurement of Surface Topography,” Advances in Optics and Photonics, 7(1), pp. 1–65. https://doi.org/10.1364/AOP.7.000001.
[8] Li, J., Ma, R., and Bai, J., 2024, “High-Precision Chromatic Confocal Technologies: A Review,” Micromachines, 15(10), 1224. https://doi.org/10.3390/mi15101224.
[9] Hu, Y., Chen, Q., Feng, S., and Zuo, C., 2020, “Microscopic Fringe Projection Profilometry: A Review,” Optics and Lasers in Engineering, 135, 106192. https://doi.org/10.1016/j.optlaseng.2020.106192.
[10] Feng, S., Zuo, C., Zhang, L., Tao, T., Hu, Y., Yin, W., Qian, J., and Chen, Q., 2021, “Calibration of fringe projection profilometry: A comparative review,” Optics and Lasers in Engineering, 143, 106622. https://doi.org/10.1016/j.optlaseng.2021.106622.
[11] Chang, B. R., Tsai, H. F., and Mo, H. Y., 2024, “Ensemble Meta-Learning-Based Robust Chipping Prediction for Wafer Dicing,” Electronics, 13(10), 1802. https://doi.org/10.3390/electronics13101802.
[12] Friedrich, M., Schlosser, T., and Kowerko, D., 2026, “Simulation of semiconductor wafer dicing induced faults on chips and their application as augmentation method for a deep learning based visual inspection system,” Journal of Intelligent Manufacturing, 37(2), pp. 573–596. https://doi.org/10.1007/s10845-024-02559-0.
[13] International Organization for Standardization, 2025, “Geometrical Product Specifications (GPS) — Surface Texture: Areal — Part 604: Design and Characteristics of Non-Contact (Coherence Scanning Interferometry) Instruments,” ISO 25178-604:2025, Geneva, Switzerland.
[14] Montgomery, P. C., Flury, M., Anstotz, F., Marbach, S., Cordier, C., Bartringer, J., Mukhtar, H., Leong-Hoï, A., Rubin, A., Shpiruk, A., Del Nero, M., and Barillon, R., 2023, “Characterization of Functional Materials Using Coherence Scanning Interferometry and Environmental Chambers,” ACS Omega, 8(12), pp. 10643–10655. https://doi.org/10.1021/acsomega.2c07007.
[15] Ma, L., Yang, F., Zhao, Y., Pei, X., and Yin, X., 2023, “Carrier phase distribution based scan step noise correction for white light interferometry topography measurements,” Optics and Lasers in Engineering, 169, 107730. https://doi.org/10.1016/j.optlaseng.2023.107730.
[16] Wu, D., Liang, F., Kang, C., and Fang, F., 2021, “Performance Analysis of Surface Reconstruction Algorithms in Vertical Scanning Interferometry Based on Coherence Envelope Detection,” Micromachines, 12(2), 164. https://doi.org/10.3390/mi12020164.
[17] Li, Y., Yang, Y., Zhang, J., Yuan, Q., and Liang, Y., 2023, “Envelope peak detection algorithm based on the CEEMDAN in white light interferometry,” Optics Continuum, 2(8), pp. 1875–1888. https://doi.org/10.1364/optcon.496459.
[18] Li, J., and Zhang, Q., 2022, “Two-step orthogonalization phase demodulation method based on a single differential interferogram,” Optics Express, 30(20), pp. 35467–35477. https://doi.org/10.1364/oe.470844.
[19] Larkin, K. G., 1996, “Efficient nonlinear algorithm for envelope detection in white light interferometry,” Journal of the Optical Society of America A, 13(4), pp. 832–843. https://doi.org/10.1364/josaa.13.000832.
[20] Kim, J., Kim, G. H., Park, S. M., Choi, E. S., Lee, S. S., Kim, K. W., Kim, H. H., Kim, C. S., and Lee, H., 2025, “Dynamic-range-extended line-field interferometric profiling with high-speed in-fiber optical delay control,” Photonics Research, 13(11), pp. 3210–3219. https://doi.org/10.1364/prj.565450.
[21] Kahatapitiya, N. S., Saleah, S. A., Seong, D., Ravichandran, N. K., Han, S., Wijesinghe, R. E., Jeon, M., and Kim, J., 2026, “Optical Coherence Tomography for High-Precision Industrial Inspection in Industry 4.0: Advances, Challenges, and Future Trends,” Laser & Photonics Reviews, 20(8), e02290. https://doi.org/10.1002/lpor.202502290.
[22] Bouchal, P., Konečný, M., Nováček, Z., Chmelík, R., and Bouchal, Z., 2026, “Coherence-Gated Mirau Interferometry for Single-Shot Quantitative Phase Imaging and Motion Sensing,” Laser & Photonics Reviews, 20(10), e02417. https://doi.org/10.1002/lpor.202502417.
[23] Zhao, H., Li, Z., Wu, Q., Wang, J., Deng, H., Liu, S., and Zhu, J., 2026, “Epi‐Illumination Spatio‐Temporal Scanning Interference Microscopy for 3D Topography Measurement With an Unlimited Field‐of‐View,” Laser & Photonics Reviews, 20(9), e02509. https://doi.org/10.1002/lpor.202502509.
[24] Li, Y., Yuan, Q., Huo, X., Wang, S., He, H., and Gao, Z., 2026, “Fourier Ptychographic Coherence Scanning Interferometry for 3D Morphology of High Aspect Ratio and Composite Micro-Trenches,” Light: Science & Applications, 15(1), 93. https://doi.org/10.1038/s41377-026-02189-6.
[25] Petzing, J., Coupland, J., and Leach, R. K., 2010, The Measurement of Rough Surface Topography Using Coherence Scanning Interferometry, Measurement Good Practice Guide No. 116, National Physical Laboratory, Teddington, UK.
[26] Chu, H. H., Kim, D. H., Park, J. H., Kang, S., Son, J., Lee, H., Yoo, H., Kim, S. W., Kim, S., and Kim, Y. J., 2025, “Nanometer-resolution white-light scanning interferometry for surface-profiling of hybrid bonding samples for advanced semiconductor packaging,” Applied Surface Science, 689, 162466. https://doi.org/10.1016/j.apsusc.2025.162466.
[27] Gao, S., Li, Z., and Brand, U., 2024, “Neural Network Approach for Modelling and Compensation of Local Surface-Tilting-Dependent Topography Measurement Errors in Coherence Scanning Interferometry,” Metrology, 4(3), pp. 446–456. https://doi.org/10.3390/metrology4030027.
[28] Liu, T., Wang, J., Liu, Q., Hu, J., Wang, Z., Wan, C., and Yang, S., 2022, “Chromatic confocal measurement method using a phase Fresnel zone plate,” Optics Express, 30(2), pp. 2390–2401. https://doi.org/10.1364/oe.449031.
[29] Liao, M., Yang, Y., Lu, X., Li, H., Zhang, J., Wang, J., and Chen, Z., 2024, “Off-Axis Co-Optical Path Large-Range Line Scanning Chromatic Confocal Sensor,” Photonic Sensors, 14(3), 240309. https://doi.org/10.1007/s13320-024-0713-5.
[30] Wang, S., Diao, K., Liu, X., and Gong, C., 2024, “Optical path optimization of chromatic line confocal displacement sensor for high resolution and wide range,” Optics Letters, 49(6), pp. 1445–1448. https://doi.org/10.1364/ol.518595.
[31] Chen, L. C., Tan, P. J., Wu, G. W., Lin, C. J., and Nguyen, D. T., 2021, “High-speed chromatic confocal microscopy using multispectral sensors for sub-micrometer-precision microscopic surface profilometry,” Measurement: Sensors, 18, 100165. https://doi.org/10.1016/j.measen.2021.100165.
[32] Gladines, J., Sels, S., Hillen, M., and Vanlanduit, S., 2022, “A Continuous Motion Shape-from-Focus Method for Geometry Measurement during 3D Printing,” Sensors, 22(24), 9805. https://doi.org/10.3390/s22249805.
[33] Weng, J. F., 2025, “Fourier-based structural metric with a CNN-assisted focus decision for robust autofocusing in microscopy,” Applied Optics, 64(35), pp. 10505–10515. https://doi.org/10.1364/ao.580406.
[34] Ashfaq, K., and Mahmood, M. T., 2024, “Enhancing Focus Volume through Perceptual Focus Factor in Shape-from-Focus,” Mathematics, 12(1), 102. https://doi.org/10.3390/math12010102.
[35] Yang, X., Zhang, H., Liu, Z., Fan, Y., Yue, S., Ma, D., Chen, X., and Wang, P., 2023, “Time‐stretch Chromatic Confocal Microscopy for Multi‐Depth Imaging,” Laser & Photonics Reviews, 17(12), 2300387. https://doi.org/10.1002/lpor.202300387.
[36] Zhang, X., Qiu, L., Zhao, W., Fu, Y., Wang, Y., and Liu, Y., 2025, “Free-form surface measurement with laser differential confocal precise positioning,” Optics & Laser Technology, 183, 112325. https://doi.org/10.1016/j.optlastec.2024.112325.
[37] Shen, S., Liu, Y., Ma, L., Ji, Z., Lin, D., and Chen, X., 2025, “Self‐Powered Perovskite Photovoltaic Photodetector Enabled Hyper‐Sensitive Reflectance Confocal Microscopy,” Laser & Photonics Reviews, 19(21), e00147. https://doi.org/10.1002/lpor.202500147.
[38] Zhang, H., Zhou, J., Jia, D., Huang, J., and Yuan, J., 2024, “Two-plus-two fringe projection profilometry based on phase-shifted coding,” Scientific Reports, 14(1), 21768. https://doi.org/10.1038/s41598-024-71427-x.
[39] Wang, D., Yan, N., Liu, H., Li, C., Yin, S., and Zhang, X., 2022, “Study on microscopic fringe structured light measurement method for tiny and complex structural components,” Applied Physics B, 128(11), 207. https://doi.org/10.1007/s00340-022-07919-3.
[40] Zeng, J., Li, Y., Li, S., and Tan, M., 2024, “Toward higher-frequency fringe projection profilometry with fewer patterns,” Optics Letters, 49(23), pp. 6649–6652. https://doi.org/10.1364/ol.542684.
[41] Wang, B., Chen, W., Qian, J., Feng, S., Chen, Q., and Zuo, C., 2025, “Single-Shot Super-Resolved Fringe Projection Profilometry (SSSR-FPP): 100,000 Frames-per-Second 3D Imaging with Deep Learning,” Light: Science & Applications, 14(1), 70. https://doi.org/10.1038/s41377-024-01721-w.
[42] Lv, S., and Kemao, Q., 2023, “Modeling the measurement precision of Fringe Projection Profilometry,” Light: Science & Applications, 12(1), 257. https://doi.org/10.1038/s41377-023-01294-0.
[43] Wang, M., Hou, Q., Yin, Y., Yang, Y., Tang, Q., Peng, X., and Liu, X., 2022, “Large DOF microscopic fringe projection profilometry with a coaxial light-field structure,” Optics Express, 30(5), pp. 8015–8026. https://doi.org/10.1364/oe.452361.
[44] Zhang, G., Liu, Y., Yao, Q., Deng, H., Zhao, H., Zhang, Z., and Yang, S., 2024, “Multi-view fringe projection profilometry for surfaces with intricate structures and high dynamic range,” Optics Express, 32(11), pp. 19146–19162. https://doi.org/10.1364/oe.524898.
[45] Wang, D., Zhou, W., Zhang, Z., Kang, Y., Meng, F., and Wang, N., 2023, “Anti-crosstalk absolute phase retrieval method for microscopic fringe projection profilometry using temporal frequency-division multiplexing,” Optics Express, 31(24), pp. 39528–39545. https://doi.org/10.1364/oe.506370.
[46] Chen, S., Li, Y., and Zhang, G., 2025, “OpticFusion: Multi-Modal Neural Implicit 3D Reconstruction of Microstructures by Fusing White Light Interferometry and Optical Microscopy,” 2025 International Conference on 3D Vision (3DV), pp. 90–100. https://doi.org/10.1109/3dv66043.2025.00014.
[47] Ye, R., Zeng, J., Zhang, H., Su, Y., and Li, H., 2025, “Rapid Full-Field Surface Topography Measurement of Large-Scale Wafers Using Interferometric Imaging,” Photonics, 12(9), 835. https://doi.org/10.3390/photonics12090835.
[48] Hu, J., Liu, T., Ji, P., Zhang, G., Huang, Y., Cheung, C. F., and Yang, S., 2025, “High-sensitivity, high-throughput inspection of nanoscale defects using a laser confocal positioning-assisted darkfield imaging system,” Optics & Laser Technology, 191, 113269. https://doi.org/10.1016/j.optlastec.2025.113269.
[49] Choi, H., Kwak, H., Ryu, S., Sohn, Y., and Kim, J., 2025, “Backside illumination-enabled metrology and inspection inside 3D-ICs using frequency comb-based chromatic confocal and spectral interferometry,” Light: Advanced Manufacturing, 6(4), pp. 791–804, Article 74. https://doi.org/10.37188/lam.2025.074.
[50] Zhang, R., Guo, X., Xu, Y., Zhang, T., and Ming, M., 2026, “BWD-DETR: A Robust Framework for Bright-Field Wafer Defect Detection,” Sensors, 26(3), 1064. https://doi.org/10.3390/s26031064.
[51] Park, J., Choi, Y., Kwon, S., Lee, Y., Kim, J., Kim, J.-J., Lee, J., Ahn, J., Kwak, H., Yang, Y., Jo, T., Lee, M., and Kim, K., 2024, “Microsphere-assisted hyperspectral imaging: super-resolution, non-destructive metrology for semiconductor devices,” Light: Science & Applications, 13(1), 122. https://doi.org/10.1038/s41377-024-01469-3.
[52] Gollee, C., Seidel, A., Schenke, C.-C., Hellmich, A., and Ihlenfeldt, S., 2022, “Development and Extrinsic Calibration of a 3D Optical Multisensor Platform Using Laser Line Scanner and a Three-Axis Linear Motion Unit,” ASME J. Manuf. Sci. Eng., 144(9), 095001. https://doi.org/10.1115/1.4054240.